98648-006-06 Sartorius IS 150 IGG-H, IS 150 IGG-H0CE IS 300 IGG-H, IS 300IGG-H0CE Weighing Platforms Installation and Operating Instructions
10 The following diagrams show the options for powering weighing platforms and industrial terminals. Cables for connecting the industrial-grade AC ad
11 Safety Precautions The weighing platform is energized by an external industrial-grade AC adapter. Make sure that the voltage rating printed on thi
12Whenever possible, avoid wiring your weighing platform into networks that carry a heavy electrical load (for instance, a compressor or similar equip
13 Operating Limits Maximum Overload Capacity Sartorius weighing platforms are built so that occasionally loading them beyond their maximum weighi
14 General Instructions for Integration into Conveyor Systems The IS weighing platform is suitable for installation in conveyor systems. Follow the g
15 15 Dimensions in millimeters
16 Pit Installation of the Weighing Platform Preparation Choose the appropriate pit frame from the list of accessories. Note to Users in the EU* I
17 Aufstellen des Rahmens für Grubeneinbau – Schrauben (7), Scheiben und Metallplatten am Grubenrahmen entfernen – Gewinde des Grubenrahmens abkle
18 Installation of the Drive-on Ramp YAR 10 IS To install the drive-on ramp, YAR 10 IS, use the angular braces and screws provided. Proceed as sho
19 Configuring the Weighing Platform You can adapt the weighing platform to meet your specific requirements – e.g., metrological specifications, f
2 Contents Page General View of the Weighing Platform 3 Warranty 4 Storage and Shipping Conditions 4 Safety Instructions 5 Installation
20 Specifications Charts Access the configuration menu of the conected Sartorius terminal to set the code for the group of specifications of your cho
21 IS 300 IGG-H0CE Weighing range structure SuperRange SingleRange Multi-Interval Weighing range Standard Configurable Maximum capacity, Max kg
22General Specifications (for Use in Standard Applications Not Subject to Legal Metrology) Model IS 150 IGG-H IS 300 IGG-H Readability g 1 2 Stand
Selecting and Attaching ID Plates The ID plate is to be selected and attached according to the specification block (configuration of scales) and the
24 Pin Assignment Chart for the Interface RS-485 RS-485, 3-m Cable Male Connector: Type C091D, 12-pin round connector with threaded locking ring, Amph
25 Troubleshooting Guide (For Problems That You Can Fix Yourself) Problem Possible Causes Solution No segments appear on the display – No line
26 Problem Possible Causes Solution The error code "ERR 12" is displayed – You tried to store a value in the tare memory for applicati
27 Care and Maintenance Servicing Regular servicing by a Sartorius technician will extend the service life of your weighing platform and ensure it
28 Safety Inspection If there is any indication that safe operation of the weighing platform is no longer warranted, turn off the power and discon
3 General View of the Weighing Platform No. Designation 1 Load plate 2 Manufacturer's label 3 Line voltage connector (optional) 4 Leveli
30
32
35
36 Use of the Weighing Platform as a Legal Measuring Instrument in the EU* Components of a Legal Weighing The weighing platform is a modular devic
37 Effect of the Preload on the Acceptability for Verification in the EU* Only internal adjustment (calibration) is allowed for weighing instrumen
38 Approved Auxiliary Measuring Devices When using the weighing platform as a legal measuring instrument in the EU, you may connect to it only auxil
4 With this Sartorius IS weighing platform, you have acquired a high-quality weighing instrument that features advanced technology. As a rule, you
41 Declarations of Conformity The C Mark on Sartorius Equipment In 1985, the Council of the European Community approved a resolution concerni
42 Additional Directive for Weighing Instruments Used in Legal Metrology: Council Directive 90/384/EEC "Non-Automatic Weighing Instruments"
43
44
45
47 "EC Verification" - A Service Offered by Sartorius Our service technicians au
48 Accessories (Options) Order No.: Data interface port with 3-m cable (IP67, RS-485) and 12-pin round connector YDO50IS Data interface port wit
49 PC adapter cable, 25-pin D-Submini – 9-contact 6965619 D-Submini 0.25 m long (YDO 52 IS and YCC 01-0017D2 required) S
5 Safety Instructions ! The seals affixed to this equipment indicate that only authorized service technicians are allowed to open the equipment and
50 Roller conveyor, with 8 rollers (for the 800 x 600mm IS weighing platform) Aluminum YRC01IGP Stainless steel YRC01IGS Drive-on ram
51
Sartorius Weighing Technology GmbHWeender Landstraße 94-10837075 Goettingen, GermanyPhone +49.551.308.0Fax +49.551.308.3289www.sartorius-mechatronics.
6 Installation Instructions Sartorius dealers or service technicians, who have received special training, will help you set up the weighing platform
7 Getting Started Unpacking the Weighing Platform Important Note To install the weighing platform in a pit, please observe the special requirement
8 Leveling the Weighing Platform Using the Level Indicator At the place of installation, level the weighing platform using the leveling fe
9 Connecting the Weighing Platform Before initially operating the scale, connect the following cables and completely assemble the com