Sartorius CPA12001S-OCE Specifications Page 40

  • Download
  • Add to my manuals
  • Print
  • Page
    / 88
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 39
Specifications
MA35|MA100|MA150
MA35 MA100 MA150
Max. weighing capacity (g) 35 100 150
Accuracy of the weighing system (mg) 1 0.1 1
Weighing system with EC type-approval
certificate
Repeatability, average (%)
– for initial sample weight approx. >1 g ± 0.2 ± 0.1 ± 0.2
– for initial sample weight approx. >5 g ± 0.05 ± 0.02 ± 0.05
Readability (%) 0.01 0.001 0.01
Display mode for results
– % moisture
– % dry weight
– % RATIO
– g residue
– g/kg residue
– g/l residue
– mg weight loss
– Calculated value (measured value + factor) •
Temperature range and settings
– 40°C–160°C, adjustable in 1-degree increments
– 30°C–230°C, adjustable in 1-degree increments
– 40°C–220°C, adjustable in 1-degree increments
Heating mode
– Standard drying
– Quick drying
– Gentle drying
– Phase drying 3+0.1–999 min. 1+0.1–999 min.
Analysis mode
– Fully automatic
– Semi-automatic 1–50 mg/5–300 sec. 1–50 mg/5–300 sec.
0.1–5.0%/5–300 sec. 0.1–5.0%/5–300 sec.
– Timer settings 1+0.1–99 min. 3+0.1–999 min. 1+0.1–99 min.
– Timer mode + fully/semi-automatic 2+0.1–999 min. +
automatic
SPRM
®
mode for parameter recognition
Heating unit
– Ceramic IR heating element (infrared)
– Halogen lamp (infrared)
– CQR heater (coiled quartz radiator)
– Metal tubular-shaped heating element
(infrared dark radiator)
Later exchange of the heating unit
by Plug & Dry
®
*
Access to the sample chamber
– via hinged, flip-up cover
– via motorized cover
40