Conversion (Secura® and Quintix® only)
User Manual for Secura, Quintix, Practum
Conversion (Secura
®
and Quintix
®
only)
Purpose: Use this application to multiply the weight value by a user-defined factor. Division
is also possible if a factor of less than zero is selected. The selected factor is saved
to protected memory.
Example: You want to calculate the weight per unit of a sheet of paper (DIN A4).
To do so, proceed as follows:
The weight per unit (e.g., 80 g/m
2
) can be determined by dividing the weight
by the surface.
The surface of a sheet of DIN A4 paper is:
0.210 0.297 = 0.06237 m
2
.
Division by 0.06237 is equal to multiplication by reciprocal value 1 / 0.06237,
i.e., 16.03335.
Set the factor to 16.03335 in the application.
t Select the Menu key in any application.
t Select the Conversion symbol in the menu.
Max 2100 g d
=
0.01 g
0.00
Factor
1
y The Conversion application appears.
The multiplier is displayed under Factor.
Factor
1
t To set the factor and accuracy of the conversion, select the gray button.
Conversion
Factor Accuracy
16.03335
1
2
0
0.0
0.00
y The Conversion settings window appears.
t On the left, enter the factor by which the weight of the sample will be
multiplied.
Select a value or select t, enter the desired value, and confirm with l.
t Select the accuracy of the conversion display on the right.
t Select l to confirm.
t Select A to zero the balance if necessary.
t Select w.