57
Unstable Condition
User Manual for Secura, Quintix, Practum
Unstable Condition
Purpose: Use this application with moving samples (such as live animals) and for weighing
in unstable environments. A measurement cycle is automatically carried out with
a defined number of measurements for each object to be weighed. The individual
measurements are averaged, and this average is displayed as the result.
t Select the Menu key in any application.
t Select the Unstable condition symbol in the menu.
Max 2100 g d
=
0.01 g
0.00
Measurements
5
y The Unstable condition application appears.
The set number of measurements is displayed under
Measurements.
Measurements
5
t To set the number of measurements and other settings, select the gray button.
Unstable condition
Measurements Start
5
10
20
Instability
Auto
Manual
y The Unstable condition settings window appears.
t On the left, enter the number of measurements.
Select a value or select t, enter the desired value, and confirm with l.
t Select the degree of instability of movement in the middle (e.g., for the starting
criterion of the measurement).
t On the right, choose whether the measurements are started manually
with Manual or automatically with Auto.
t Select l to confirm.
t Select A to zero the balance if necessary.
t To weigh a living animal, place a container or cage on the weighing pan.
t Select D to tare.