Calibration and Adjustment
User Manual for Secura, Quintix, Practum
Calibration and Adjustment
Background During calibration, a calibration weight is used to determine how much the
displayed value deviates from the actual measurement value. This deviation is
compared to a specified desired value. Adjustment eliminates this deviation or
reduces the permissible error limits.
For Secura
®
, Quintix
®
, and Practum
®
models, calibration and adjustment are
combined as one process. The balance is automatically adjusted after each
calibration.
Before using a verified balance for legal metrology, calibration/adjustment must
always be carried out at the balance placement station. This can be performed
automatically or manually. Practum
®
models with accuracy class (I) can be
calibrated using an external weight.
When and How Often To achieve the highest accuracy possible, regularly calibrate and adjust the balance:
daily after switching on the balance,
each time the balance is leveled,
each time ambient conditions have changed (temperature, humidity, or air
pressure),
each time the balance is set up at a new placement station or moved on its
current station.
The following describes in detail the available options for calibrating and adjusting
the balance:
Calibration/Adjustment using an internal calibration weight (Secura
®
and
Quintix
®
only)
Calibration/Adjustment using an external calibration weight
Automatic calibration/adjustment with isoCAL